(unit 2) Which device is used to evaluate the degree of focal spot blooming ?

  • Wisconsin Test Cassette
  • Pinhole camera
  • Digital dosimeter
  • Scintillation tube - ANSWER-b. Pinhole camera

(unit 2) The focal spot size should be determined on acceptance

testing and then evaluated:

  • Monthly
  • Quarterly
  • Semiannually
  • annually - ANSWER-d. annually 1 / 4
  • | P a g e

(unit 2) The light field and the radiation field must be congruent

to within:

  • +/-2% of the SID
  • +/-5% of the SID
  • +/-7% of the SID
  • +/-10% of the SID - ANSWER-a. +/-2% of the SID

(unit 2) Image receptor-radiation field alignment can be

evaluated using:

  • A star resolution pattern
  • A pinhole camera
  • The eight-penny test
  • The Wisconsin test cassette - ANSWER-c. The eight-penny
  • test

(unit 2) Another term for automatic collimation is:

  • Positive beam limitation
  • Scintillation
  • Congruence
  • perpendicularity - ANSWER-a. Positive beam limitation 2 / 4
  • | P a g e

(unit 2) The illumination of the x-ray tube light source at 100cm

must be at least:

  • 40 lux
  • 80 lux
  • 120 lux
  • 160 lux - ANSWER-d. 160 lux

(unit 2) The x-ray tube must be mounted in its housing so that the central ray of the x-ray beam is within ___ degree of perpendicular.

  • 1
  • 2
  • 3
  • 4 - ANSWER-a. 1

(unit 2) The overload protection mechanism of an x-ray generator should not permit an exposure that exceeds ___ of the tube capacity for a single exposure.

  • 50%
  • 60% 3 / 4
  • | P a g e
  • 70%
  • 80% - ANSWER-d. 80%

(unit 2) The most common type of sensor found in AEC systems

is the:

  • Photodetector
  • Ion chamber
  • Solid-state detector
  • photocell - ANSWER-b. Ion chamber

(unit 2) It is estimated that more than ___ of all hospital radiology departments have radiographic equipment that uses an AEC system.

  • 30%
  • 50%
  • 60%
  • 90% - ANSWER-c. 60%

(unit 2) The backup timer should terminate the exposure within

6 seconds or:

  • 200 mAs
  • / 4

Download Document

Buy This Document

$30.00 One-time purchase
Buy Now
  • Full access to this document
  • Download anytime
  • No expiration

Document Information

Category: Business
Added: Aug 7, 2025
Description:

CALIFORNIA SUPERVISOR AND OPERATOR ASRT NEWEST 2025 ACTUAL EXAM COMPLETE 200 QUESTIONS AND CORRECT DETAILED ANSWERS (VERIFIED ANSWERS) |ALREADY GRADED A+ (unit 2) Which device is used to evaluate t...

Get this document $30.00