DaVita star learning

EXAM ELABORATIONS Aug 29, 2025
Loading...

Loading document viewer...

Page 0 of 0

Document Text

DaVita star learning The primary electrolyte in both our blood and the dialysate is... - correct answer sodium Policy states the conductivity meter should be calibrated when an inaccurate reading is suspected. It

should also be calibrated: - correct answer prior to daily use

When the dialysate sodium level is higher than the level in the blood: - correct answer Diffusion will cause sodium to move into the blood The calibration of the conductivity meter should be completed at least... - correct answer daily Hypernatremia can lead to... - correct answer -Excessive thirst Decreased weight gains between treatments Comparing the displayed machine conductivity and your manual conductivity reading with the Fresenius machine's TCD reading helps to... - correct answer -Assure the machine reading is accurate

  • Identify malfunctions of the machine's proportioning system
  • Identify malfunctions of the machine's alarm/bypass system
  • When comparing a manual conductivity reading to the applicable machine conductivity reading, the difference in readings should be no greater than.. - correct answer 0.4 mS/cm If, after recalibrating the conductivity meter and retesting the machine, the difference between the manual and applicable machine conductivity remains greater than 0.4 mS/cm, one should.... - correct answer Remove the machine, tag the machine, and notify biomed. 1 / 2

True/false: A manual conductivity reading should not be completed when the machine is in bypass mode. - correct answer true

True/false: The primary reason for comparing a

manual conductivity with the applicable machine conductivity is to assure patient safety. - correct answer true Machine alarm tests are performed: - correct answer Before the initiation of each patient treatment Once set up for treatment, the recirculation time for a dialyzer should not exceed - correct answer 2 hours The manual conductivity must match the Fresenius (FMC) dialysis delivery system displayed conductivity

  • correct answer +/- 0.4mS
  • To prepare a 1:100 bleach solution, you would - correct answer Add 1 part of bleach to 99 parts of water What happens if bleach and peracetic acid are mixed? - correct answer A chemical reaction occurs, producing toxic chlorine gas What is the correct procedure for residual bleach testing after chemical disinfection of distribution loops and equipment fluid pathways? - correct answer Residual bleach testing is performed after bleach disinfection and prior to use What is the safe range for dialysate pH? - correct answer Between 6.9 and 7.6 Why is a "strikethrough"‖ of an external pressure transducer protector of concern? - correct answer Fluid could have entered the machine and contaminated the internal pressure transducer protector providing a reservoir for microorganisms and causing subsequent patient blood infections

  • / 2

Download Document

Buy This Document

$30.00 One-time purchase
Buy Now
  • Full access to this document
  • Download anytime
  • No expiration

Document Information

Category: EXAM ELABORATIONS
Added: Aug 29, 2025
Description:

DaVita star learning The primary electrolyte in both our blood and the dialysate is... - correct answer sodium Policy states the conductivity meter should be calibrated when an inaccurate reading i...

Get this document $30.00