IBHRE EXAM
What is meant by "intention to treat" analysis in a randomized control trial?
- patients are analyzed in the groups they were randomized to
- patients are analyzed in the groups they were randomized to
- patients are analyzed in the groups they were randomized to
- patients are only included in the analysis whose treatment was
regardless of whether they complied with the treatment given
only if they actually received treatment
unless they crossed-over into another treatment arm
at least somewhat effective - CORRECT ANSWER -A. Patients Are Analyzed In The Groups They Were Randomized To Regardless Of Whether They Complied With The Treatment Given
Which of the following formulas represents the fundamental unit of work? 1 / 4
- | P a g e
a) E=I2 x R x T
b) E=V x I x T
c) E=V2/R x T
- all of the above - CORRECT ANSWER -D. All the above
The time from electrical depolarization of the heart to mechanical systole describes which of the following terms?
- evoked response interval
- pre-ejection interval
c) QT interval
- dV/dT - CORRECT ANSWER -B. Pre-ejection Interval
Typical ventricular blanking periods in DDD pacemakers range
between:
- 1 to 10 milliseconds
- 12 to 50 milliseconds
- 50 to 100 milliseconds
- 100 to 200 milliseconds - CORRECT ANSWER -B. 12 to 50
ms 2 / 4
- | P a g e
The basic unit of capacitance is:
- joule
- farad
- henry
- coulomb - CORRECT ANSWER -B. Farad
Which of the following sources of device interference would be most likely to cause rapid pacing or reset to back-up mode?
- cell phone
b) MRI
- electronic article surveillance
- airport metal detectors - CORRECT ANSWER -B. MRI
The reported effects on pacemakers of radiofrequency catheter ablation performed in close proximity (< 4 cm) to pacing lead
systems include(s):
- induction of pacemaker runaway
- pacemaker reversion to noise mode behavior 3 / 4
- | P a g e
- inappropriate inhibition of pacemaker
- all of the above - CORRECT ANSWER -D. All the above
Which of the following would increase pacing lead impedance?
- smaller electrode
- shorter conductor
- insulation break
- programming to unipolar - CORRECT ANSWER -A. Smaller
Electrode
Polarization impedance refers to what general area of the pacing system?
- electrode/tissue interface
- lead body
- header
- circuit board - CORRECT ANSWER -A. Electrode / Tissue
Interface
A pause in paced rhythm below the base rate may be normally
explained by: