MDCB: Mock Exam 2 Questions with 100%
Verified Correct Answers A patient receiving treatment to the head and neck region has experienced a skin reaction and mucositis at a total dose much lower than anticipated. The radiation oncologist requests an evaluation of the treatment. Which of the following steps should be taken?
- recalculate the treatment plan
- ensure that the depth dose, build-up region, wedge parameters, and other pertinent information
- review the treatment setup with the therapist
- recalculate the monitor unit settings
- all are correct - Correct Answer E
for the treatment unit are correct
For megavoltage dosimetry, TLD can provide accuracy of:
- plus/minus 20%
- plus/minus 10%
- plus/minus 3%
- plus/minus 1% - Correct Answer C
With megavoltage film dosimetry, isodose curves can be measured to within:
- plus/minus 10%
- plus/minus 7% 1 / 4
- plus/minus 3%
- plus/minus 1% - Correct Answer C
In megavoltage therapy, typical use(s) of diodes is for:
- patient dosimetry
- beam scanning
- quality assurance
- all of the answers are correct - Correct Answer D
Which of the following is best suited for calibration of a megavoltage beam:
- diode detector
- thermoluminescence dosimetry
- film
- ionization chamber - Correct Answer D
The f-factor:
- is the roentgen to cGy conversion factor
- is greatest for high Z materials and low energy photons 2 / 4
- is 0.873 in air
- all of the answers are correct
- none of the answers are correct - Correct Answer D
- shaping of electron fields as a replacement for cerrobend blocks
- dynamic field shaping as a function of patient respiration
- field shaping without penumbra
- modulating beam intensity - Correct Answer D
Clinical application of MLC include which of the following?
The 104% isodose lines at the four corners of the target shown are due to the:
- inhomogeneity of the femur
- scatter from the femur
- horns of the beams
- weights of the lateral beams
- use of a wrong algorithm - Correct Answer C
Exposure is:
- the amount of energy in Joules/kg transferred from a photon beam to a material 3 / 4
- only defined for photons and charged particles below 3 MeV
- the charge liberated by photons in a given mass of air
- the absorbed dose multiplied by the quality factor
- None of the answers - Correct Answer C
The quantity that an ionization chamber actually measures is:
- roentgen
- gray
- charge
- kerma
- voltage - Correct Answer C
Geiger counters:
- are sensitive enough to detect individual photons or particles
- are less sensitive than ionization chambers
- are used to calibrate linac output
- operate on the principle of scintillation - Correct Answer A
- / 4