Question : PDD is dependent on the following factors
Correct answer: Increases with beam energy
Increases with field size Increases with SSD Decreases with Depth
Question : TAR is dependent on the following factors
Correct answer: Increases with beam energy
Increases with field size Decreases with depth Independent of SSD
Question : Skin dose is dependent on the following factors
Correct answer: Decreases with beam energy
Increases with SSD Increases with field size Increases with bolus Increases with oblique incidence
Question : The indicated field size on a medical LINAC
represents the field length and width in centimeters as defined
at the:
Correct answer: At isocenter (100 cm)
Question : On a conventional simulator, localizing lasers and
ODI checks should be performed when and within what tolerance
Correct answer: Daily, 2mm
Question : What is the tolerance for light/radiation field
coincidence?